 The TC-08 thermocouple data logger offers industry-leading performance and a cost-effective solution to your temperature measuring needs.
  • 8 channel thermocouple data logger
  • Measures from -270 to 1820 °C
  • Automatic cold junction compensation
  • High resolution and accuracy
  • Fast sampling rate
  • Multiple units can be run on a single PC
  • USB interface


screenshot

Wide temperature range

The TC-08 thermocouple data logger is designed to measure a wide range of temperatures using any thermocouple that has a miniature size thermocouple connector. Additionally, the TC-08 can also measure other sensors using a 70 mV range.

Featuring built-in cold junction compensation (CJC), the TC-08 has an effective temperature range of -270 to 1820 °C (The actual temperature range depends on the thermocouple being used.)

All the benefits of USB

The TC-08 connects to the USB port of a Windows based PC and enables the host PC to automatically detect the TC-08 — avoiding the need for any complex setup procedures. The USB connection also allows the TC-08 to be powered directly by the USB bus — eliminating the need for an external power supply and making the TC-08 ideal for measuring temperatures both in the lab and in the field.

The TC-08's USB connectivity allows up to 10 TC-08 USB data loggers to be used simultaneously on one PC — making the TC-08 ideal for use as a large multi-channel temperature acquisition system with a cost per channel significantly lower than other systems of this size.

Fast and accurate temperature data acquisition

With the TC-08 thermocouple data logger your measurements can be made both fast and accurately.

The low conversion time of the TC-08 means up to 10 temperature measurements can be taken every second. Whilst the high (20 bit) resolution ensures that the TC-08 can detect minute changes in temperature.

All you need in one easy-to-use unit

Temperature data acquisition with the TC-08 is very easy — simply plug the TC-08 into a USB port on your computer, connect your thermocouples, and you are ready to measure temperatures.

The easy-to-use PicoLog data acquisition software is supplied with the TC-08 — allowing you to start measuring and recording temperatures with the minimum amount of effort. The intuitive user interface of PicoLog makes it easy for you to set how frequently you want to take a measurement, and the number of measurements you want to take.

For a limited time, the original SERIAL TC-08 thermocouple to PC interface is still available for users without a USB compatible PC.

Specifications

Number of channels 8
Conversion time 100 ms (thermocouple and cold junction compensation)
Temperature accuracy Sum of ±0.2 % of reading and ±0.5 °C
Voltage accuracy Sum of ±0.2 % of reading and ±10 µV
Overload protection ±30 V
Maximum common mode voltage ±7.5 V
Input impedance 2 MΩ
Input range (voltage) ±70 mV
Resolution 20 bits
Noise free resolution 16.25 bits
Thermocouple types supported B, E, J, K, N, R, S, T
Input connectors Miniature thermocouple
Output connector USB — cable supplied
PC connection USB 1.1
Power supply from USB port
Dimensions 201 x 104 x 34 mm
Supplied software PicoLog data acquisition software
Driver and examples (C, Delphi and Visual Basic, Labview, Agilent VEE6.1 and Excel)
Software is supplied on CD and is compatible with Windows 98SE/ME/2000/XP
Price +- $459.00 (US) See Price list for exact price

TC-08 Resolution

Thermocouple type Overall range °C 0.1°C resolution 0.025°C resolution
B 20 to 1820 150 to 1820 600 to 1820
E -270 to 910 -270 to 910 -260 to 910
J -210 to 1200 -210 to 1200 -210 to 1200
K -270 to 1370 -270 to 1370 -250 to 1370
N -270 to 1300 -260 to 1300 -230 to 1300
R -50 to 1760 -50 to 1760 20 to 1760
S -50 to 1760 -50 to 1760 20 to 1760
T -270 to 400 -270 to 400 -250 to 400

Applications for the TC-08

The TC-08 thermocouple data logger is used successfully throughout the industrial, scientific, medical and the food industries; listed here are some interesting applications:

  • Generation of electricity from biomass — an investigation by the UK government into renewable energy sources
  • Espresso coffee grinder testing — how the TC-08 is helping to make the perfect Espresso
  • See our thermocouples application note

Image of the SE000, SE027, SE028 & SE029 series of thermocouples:

 

Put Links and images as needed here

Return to top of page


 USA Office: Tel: 1-877-902-2979    Fax: 1-877-FAX-IECI     Pacific  Time Zone    
                   Tel:1-425-223-4311    Fax: 1-425-223-4303     Pacific Time Zone
P.O. Box 1280. Street Address: 145 Tyee Drive, Point Roberts, WA 98281

Sales inquiries to:  sales at interworldusa.com  Support & Service: support at interworldusa.com

Western Canada: - Vancouver, B.C. Tel: 1-(604) 925-6150    Fax: 1-(604) 925-6170 Toll Free:1-(800) 663-6001

    2454 Haywood Ave, West Vancouver, B.C. V7V 1Y1
  Eastern Canada: - Markham, Ontario Tel: 1-(905) 513-7027    Fax: 1-(905) 513-7029 Toll Free:1-(800) 465-0164
3555 14th Ave, Unit 12, Markham, Ontario L3R 0H5

Sales inquiries to:  sales at interworld.ca    Support & Service: support at interworld.ca

To visit our Australian web site, please click here.